> For the complete documentation index, see [llms.txt](https://mpqc.swcmicroscopy.com/llms.txt). Markdown versions of documentation pages are available by appending `.md` to page URLs; this page is available as [Markdown](https://mpqc.swcmicroscopy.com/protocols/protocol-full-characterisation.md). Agents can also search and read this documentation through its [MCP server](https://mpqc.swcmicroscopy.com/~gitbook/mcp).

# Protocol - Full characterisation

This protocol should be performed for a full characterisation of the microscope setup. It will provide baseline data for quality control checks and can be used to ensure accurate reporting of microscope characteristics in publications.

#### Back up the machine data file

&#x20;Run:

```
mpqc.record.MDFb
```

#### Laser power

Place a powermeter under the objective. Run:

```
mpqc.record.power
```

Turn on the laser then wait half an hour before starting. If you use multiple wavelengths, record the power curve for most commoly used wavelengths. If there is a very large change in power compared to last time you should investigate, check if the beam being clipped or if the laser needs a service.

1. Zero the sensor under the objective&#x20;
2. Measure power curve&#x20;
3. Save data&#x20;
4. (if required) Press calibrate power button&#x20;

\*\* power curve must first be generated using native scanimage calibration software

#### Electrical noise

Remove anything from under the objective. Run:

```
mpqc.record.electrical_noise
```

#### Lens paper

Find the focus of the lens paper. For information on how to create the lens paper slide, see materials section. For photon counting measurements, image the lens paper at 100mW. For low signal imaging, image at 20mW or less.

```
mpqc.record.lens_paper('power',100,'wavelength',800)
```

#### Uniform slide

Place a uniform green slide under the objective. Find the surface and go down until the image no longer appears to change as you focus down. i.e. try to image as close to the surface as possible bearing in mind that your field is probably curved. Image at low power and use only a single detector. Use the green PMT at low gain unless the signal is too strong, in that case use the red PMT. Run:

```
mpqc.record.uniform_slide('power',5,'wavelength',920)
```

#### Standard light source

Place a standard light source under the objective. For consistent recording move the objective as close as possible to the source. Run:

```
mpqc.record.standard_light_source
```

For further details on making a standard light source, see materials section.

#### EM grid

Place EM grid slide under the objective. Find the surface of the grid and orient it such that the grid lines run vertical and horizontal, stop the acquisition. Run:

```
Grid2MicsPerPixel
```

#### PSF

Place a bead slide under the objective and locate a bead. If using the tetraspeck slide, it is easiest to find the largest bead size and then move to the smaller beads.  You might want to use 15 mW or so for finding the beads then switch to 10 mW for imaging. Image with a FOV of about 80 microns at 512 by 512 pixels. That is a zoom of around 20 (+/- 5). Use 0.25 or 0.5 microns step size in Z. Usually 15 to 20 microns of imaged thickness is enough. Average at least 8 or 16 frames per optical plane with an 8 kHz reso scanner (just set Frame Rolling Average in ScanImage to the desired value). Going up to about 40 frames per slice will yield nicer images if you care more about shape in x/y. For Z you are probably OK at the lower averaging values. Data are saved to Desktop. Make sure you know the bead diamater. If in doubt, do not use the mixed size well. Run:

```
mpqc.record.PSF
```

Tips for finding beads:

* Place slide under objective with no water. Look from the front and side to ensure bead well is under objective. Then add water.
* Set power to 30 mW, zoom 2 or 3, average 2 to 4 frames.
* If it's helpful to have higher FPS, use 256 by 256 for searching.
* Look both red and green channels.
* Usually there is a global increase in background when you are in the bead plane.
* Zero the z motor when you find the beads or think you are getting there. This stop you from getting lost.
* Set power to 10 mW before zooming in or you will melt the larger beads. Molten beads will start to slowly expand, which will mean you can no longer image them. Often they will develop strange shapes.

Once you have PSFs, use `measurePSF` to check them out. Don't proceed until you have decent images.

#### Generate PDF report

Create a report containing the characterisation of the microscope setup. Run:&#x20;

```
generateMPQCreport
```